Mitigation of photoneutron contamination in high-energy LINACs via component reconfiguration and advanced shielding

Document Type : Research Article

Authors

Faculty of Basic Sciences, Physics Department, Imam Khomeini International University, Qazvin, Iran

Abstract
Photoneutrons are produced when photons with energies greater than 10 MeV interact with high-Z materials in the accelerator head. This study investigated reducing photoneutron dose by modifying linear accelerator component configurations and using shielding materials, including Cerrobend blocks and borated polyethylene (BPE). We used the MCNPX Monte Carlo code to calculate photon and neutron fluence and dose distributions in a phantom. We evaluated different BPE and Cerrobend thicknesses for their dose-attenuation capabilities. We also examined various accelerator component configurations and calculated neutron fluence and equivalent dose using the F5 tally and dose conversion factors (DFs). Cerrobend placed around the tumor significantly reduced photon dose rates but did not substantially reduce neutron dose. Relocating the multi-leaf collimator (MLC) from below the jaws to below the primary collimator (PC) reduced neutron dose by 44%. Replacing the PC material from tungsten to iron produced a further 10% reduction in neutron dose. Using secondary collimator jaws was associated with lower neutron contamination, whereas raising the MLC increased neutron dose. The greatest neutron dose reduction was achieved with 2-4 cm thick BPE. These findings indicate that strategic modifications to accelerator design, combined with appropriate shielding materials, can effectively reduce neutron contamination during high-energy photon radiotherapy.

Highlights

  • Increasing Cerrobend thickness from 8 to 10 cm reduced neutron dose by 23%.
  • Increasing BPE thickness from 2 to 4 cm reduced neutron dose by 18%.
  • Moving the MLC beneath the primary collimator reduced neutron dose by 44%.
  • Replacing tungsten with iron in the primary collimator reduced neutron dose by 10%.
  • Flattening-filter removal had no significant effect on neutron dose

Keywords

Subjects

Copyright
RPE is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License (CC BY-NC 4.0).

Conflict of Interest
The authors declare no potential conflict of interest regarding the publication of this work‎.

Funding
‎The authors declare that no funds‎, ‎grants‎, ‎or other financial support were received during the preparation of this manuscript‎.


Articles in Press, Accepted Manuscript
Available Online from 20 November 2026

  • Receive Date 20 July 2026
  • Revise Date 26 September 2026
  • Accept Date 14 November 2026